Description
Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.
CVSS 1 Total
Score | Severity | Version | Vector String |
---|---|---|---|
7.2 | HIGH | 3.1 | CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H |
Product Status
Learn moreVersions 2 Total
Default Status: unknown
affected
Versions 2 Total
Default Status: unknown
affected
Versions 2 Total
Default Status: unknown
affected
Credits
- Julian Suleder, Nils Emmerich, and Birk Kauer of ERNW Research GmbH; Dr. Oliver Matula of ERNW Enno Rey Netzwerke GmbH, reported these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices).